TFA can be displaced from the ion-pair complex by exposing the effluent to a higher concentration of a different acid post column and before the MS detector. This is referred to as the "TFA-Fix".That is to say: Signal Suppression due to Additives (ESI signal suppression by TFA)
Ion pairing with analyte --> surface tension effect
Solutions:
Post-column addition of a sheath liquid of propanoic acid (10%) in 2-propanol (TFA Fix)
Use low concentrations of TFA with acetic acid (TFA Light)
Replace TFA
Always remember: Achievement of gaseous analyte ionization at API interface is the key to MS detection.
Reference: Agilent LC/MS Method Development Resource.
Thursday, July 24, 2008
APCI – Typical Operating Conditions
( Please look up Blog@July-06-2008 for the selecting of ionization methods for LC-MS)
Flow Rates: 50µL/min. - 2mL/min.
Vaporizer Temp (°C): 400-550 (600 max)
Discharge Current (µA): 5 (20µA max.)
Sheath Gas Flow Rate (arb): 35-80
Auxiliary Gas: 0
Capillary Temp (°C): 250-350C
Tube Lens Offset (V): 30-60V
Ensure that the APCI probe is hot enough so that the spray shield is not dripping wet.
(Note: For me, always assume that works better at higher flow rates, higher flow rate means more solvent for plasma production. Position of corona discharge needle is critical for sensitivity.)
Acids – Negative Ion detection, adjust pH 2 units below pKa.
Decrease pH use formic acid , acetic acid, TFA.
Bases – Positive Ion detection, adjust pH 2 units above pKa.
Increase pH with NH4OH, TEA, TMA.
Adjust corona discharge voltage.
Adjust nebulization temperature.
Consider possible thermal decomposition of analyte.
Flow Rates: 50µL/min. - 2mL/min.
Vaporizer Temp (°C): 400-550 (600 max)
Discharge Current (µA): 5 (20µA max.)
Sheath Gas Flow Rate (arb): 35-80
Auxiliary Gas: 0
Capillary Temp (°C): 250-350C
Tube Lens Offset (V): 30-60V
Ensure that the APCI probe is hot enough so that the spray shield is not dripping wet.
(Note: For me, always assume that works better at higher flow rates, higher flow rate means more solvent for plasma production. Position of corona discharge needle is critical for sensitivity.)
Steps for APCI Optimization
If analyte’s pKa is unknown, evaluate 3 pH regions in positive and negative ion modes.Acids – Negative Ion detection, adjust pH 2 units below pKa.
Decrease pH use formic acid , acetic acid, TFA.
Bases – Positive Ion detection, adjust pH 2 units above pKa.
Increase pH with NH4OH, TEA, TMA.
Adjust corona discharge voltage.
Adjust nebulization temperature.
Consider possible thermal decomposition of analyte.
Wednesday, July 23, 2008
Contamination Peaks - Positive Ion ES
Double check if following peaks consistently exist in the GC-MS chromatogram.
m/z Ion Compound
42 (M+H)+ Acetonitrile
59 (M+NH4)+ Acetonitrile
64 (M+Na)+ Acetonitrile
65 (2M+H)+ Methanol
79 (M+H)+ DMSO
83 (2M+H)+ Acetonitrile
85 (M+H)+ d6-DMSO
101 (M+Na)+ DMSO
102 (M+H)+ Triethylamine (TEA)
104/106 (M+Cu)+ Acetonitrile
105 (2M+Na)+ Acetonitrile
120 (M+Na+CH3CN)+ DMSO
122 (M+H)+ Tris
123 (M+H)+ Dimethylaminopyridine (DMAP)
130 (M+H)+ Diisopropylethylamine (DIPEA)
144 (M+H)+ Tripropylamine (TPA)
145/147 (2M+Cu)+ Acetonitrile
146 (3M+Na)+ Acetonitrile
150 (M+H)+ Phenyldiethylamine
153 (M+H)+ 1,8-diazabicyclo[5.4.0]undec-7-ene (DBU)
157 (2M+H)+ DMSO
159 (M+Na)+ Sodium trifluoroacetate
169 (2M+H)+ d6-DMSO
179 (2M+Na)+ DMSO
186 (M+H)+ Tributylamine
214 (M+H)+ Unidentified contaminant, maybe related to methanol
225 (M+H)+ Dicyclohexyl urea (DCU)
239/241 [(M.HCl)2-Cl]+ Triethylamine
242 M+ Tetrabutylammonium (C4H9)4N+
243 M+ Trityl Cation
257 (3M+H)+ DMSO
273 M+ Monomethoxytrityl cation (MMT)
279 (M+H)+ Dibutylphthalate (plasticiser)
301 (M+Na)+ Dibutylphthalate (plasticiser)
317 (M+K)+ Dibutylphthalate (plasticiser)
338 (M+H)+ Erucamide
360 (M+Na)+ Erucamide
391 (M+H)+ Diisooctyl phthalate (plasticiser)
413 (M+Na)+ Diisooctyl phthalate (plasticiser)
425 (M+Na)+ Unidentified contaminant (plasticiser)
429 (M+K)+ Diisooctyl phthalate (plasticiser)
449 (2M+H)+ Dicyclohexyl urea (DCU)
454 (M+Na+CH3CN)+ Diisooctyl phthalate (plasticiser)
798 (2M+NH4)+ Diisooctyl phthalate (plasticiser)
803 (2M+Na)+ Diisooctyl phthalate (plasticiser)
Thanks Dr.John Langley@University of Southampton
m/z Ion Compound
42 (M+H)+ Acetonitrile
59 (M+NH4)+ Acetonitrile
64 (M+Na)+ Acetonitrile
65 (2M+H)+ Methanol
79 (M+H)+ DMSO
83 (2M+H)+ Acetonitrile
85 (M+H)+ d6-DMSO
101 (M+Na)+ DMSO
102 (M+H)+ Triethylamine (TEA)
104/106 (M+Cu)+ Acetonitrile
105 (2M+Na)+ Acetonitrile
120 (M+Na+CH3CN)+ DMSO
122 (M+H)+ Tris
123 (M+H)+ Dimethylaminopyridine (DMAP)
130 (M+H)+ Diisopropylethylamine (DIPEA)
144 (M+H)+ Tripropylamine (TPA)
145/147 (2M+Cu)+ Acetonitrile
146 (3M+Na)+ Acetonitrile
150 (M+H)+ Phenyldiethylamine
153 (M+H)+ 1,8-diazabicyclo[5.4.0]undec-7-ene (DBU)
157 (2M+H)+ DMSO
159 (M+Na)+ Sodium trifluoroacetate
169 (2M+H)+ d6-DMSO
179 (2M+Na)+ DMSO
186 (M+H)+ Tributylamine
214 (M+H)+ Unidentified contaminant, maybe related to methanol
225 (M+H)+ Dicyclohexyl urea (DCU)
239/241 [(M.HCl)2-Cl]+ Triethylamine
242 M+ Tetrabutylammonium (C4H9)4N+
243 M+ Trityl Cation
257 (3M+H)+ DMSO
273 M+ Monomethoxytrityl cation (MMT)
279 (M+H)+ Dibutylphthalate (plasticiser)
301 (M+Na)+ Dibutylphthalate (plasticiser)
317 (M+K)+ Dibutylphthalate (plasticiser)
338 (M+H)+ Erucamide
360 (M+Na)+ Erucamide
391 (M+H)+ Diisooctyl phthalate (plasticiser)
413 (M+Na)+ Diisooctyl phthalate (plasticiser)
425 (M+Na)+ Unidentified contaminant (plasticiser)
429 (M+K)+ Diisooctyl phthalate (plasticiser)
449 (2M+H)+ Dicyclohexyl urea (DCU)
454 (M+Na+CH3CN)+ Diisooctyl phthalate (plasticiser)
798 (2M+NH4)+ Diisooctyl phthalate (plasticiser)
803 (2M+Na)+ Diisooctyl phthalate (plasticiser)
Thanks Dr.John Langley@University of Southampton
Tuesday, July 22, 2008
How to obtain a NICE GC Chromatogram with good peak profile?
To 1005ling: At the first look, seemingly found lots of impurities on the column.
(1) Did you condition the column according the GC Column manufacturer’s method? If No, Do IT now.
(2) If Yes, Double check that maybe your column was contaminated. Trim/cut the first 0.5 m of the inlet column or change another column.
If above tip doesn’t work, highly recommend you look back the Oven Temp Ramp: change the temp ramp by lowing down the initial temp, and hold for 5 min.
Or Double check the polarity different between your column and analyte.
Here is some information about the stationary phases of GC column with its applicability:
①SE-30: (100% Dimethylpolysiloxane), alternative to HP-1,HP-101, CP-si15CB,SPB-1, RTX-1,OV-1,007-1, ( similar with OV-101), is suitable for assay of amine, hydrocarbons, phenols, sulfides, flavor. Temperature Limits: 0 to 325/350°C.
② SE-52: (5%-Phenyl)-methylpolysiloxane. Or equivalent with SE-54.
③ SE-54: (5%-Phenyl)(1%-Vinyl)-methylpolysiloxane. Equivalent with HP-5,DB-5, CP-si18CB, RTX-5, OV-5, SPB-5,007-2 column. Good for assay of alkaloids, medicines, aliphatic acids, methyl ester, halide, aromatic. Temperature Limits: 0 to 350°C.
④ PEG-20M: are polymers of ethylene glycol, 20 meter. Highly cross-linked resulting in excellent peak. Alternative to HP-INNOWAX,HP-20M,CP-WAX52CB,DB-WAX,BP-20,007-CW for the assay of alcohol, aromatic, volatile oils, solvents. Temperature Limits: 0 to 250°C.
⑤ FFAP: Nitroterephthalic acid modified polyethylene glycol column. High polarity, Designed for the analysis of volatile fatty acids, ketone, nitrile and phenols. Bonded and cross-linked with solvent rinsable. Can be used at operating temperatures at 60-250ºC. Alternative to HP-FFAP, B-FFAP,CPWAX58CB,SP-1000,STABIL-PA,OV-351,007-FFAP Close equivalent to USP Phase G35.
⑥ XE-60: modified cyanoethylmethylpolysiloxane XE 60 (25% cya-. noethyl, 75% methyl)
Good for aliphatic acid,carborane, acryl amide, organophosphorus pesticides, cocaine, phenols, aromatic amine. Temperature Limits: 0 to 250°C.
⑦ OV-1701 : 14% cyanopropylphenyl-86% methyl polysiloxane. alternative to HP-1701,RTX-1701,SPB-7,SPB-1701,007-1701,good for environmental, food and beverages, pesticides, herbicides. Temperature Limits: 0 to 300 °C.
Reference: Phenomenex, Agilent, Restek Cross Reference for Similar GC Columns.
Note: Next time, pls enclose more detail of GC assay parameters, being a good questioner will help us answer well, NOT just guess. Thanks for your question, Best.
(1) Did you condition the column according the GC Column manufacturer’s method? If No, Do IT now.
(2) If Yes, Double check that maybe your column was contaminated. Trim/cut the first 0.5 m of the inlet column or change another column.
If above tip doesn’t work, highly recommend you look back the Oven Temp Ramp: change the temp ramp by lowing down the initial temp, and hold for 5 min.
Or Double check the polarity different between your column and analyte.
Here is some information about the stationary phases of GC column with its applicability:
①SE-30: (100% Dimethylpolysiloxane), alternative to HP-1,HP-101, CP-si15CB,SPB-1, RTX-1,OV-1,007-1, ( similar with OV-101), is suitable for assay of amine, hydrocarbons, phenols, sulfides, flavor. Temperature Limits: 0 to 325/350°C.
② SE-52: (5%-Phenyl)-methylpolysiloxane. Or equivalent with SE-54.
③ SE-54: (5%-Phenyl)(1%-Vinyl)-methylpolysiloxane. Equivalent with HP-5,DB-5, CP-si18CB, RTX-5, OV-5, SPB-5,007-2 column. Good for assay of alkaloids, medicines, aliphatic acids, methyl ester, halide, aromatic. Temperature Limits: 0 to 350°C.
④ PEG-20M: are polymers of ethylene glycol, 20 meter. Highly cross-linked resulting in excellent peak. Alternative to HP-INNOWAX,HP-20M,CP-WAX52CB,DB-WAX,BP-20,007-CW for the assay of alcohol, aromatic, volatile oils, solvents. Temperature Limits: 0 to 250°C.
⑤ FFAP: Nitroterephthalic acid modified polyethylene glycol column. High polarity, Designed for the analysis of volatile fatty acids, ketone, nitrile and phenols. Bonded and cross-linked with solvent rinsable. Can be used at operating temperatures at 60-250ºC. Alternative to HP-FFAP, B-FFAP,CPWAX58CB,SP-1000,STABIL-PA,OV-351,007-FFAP Close equivalent to USP Phase G35.
⑥ XE-60: modified cyanoethylmethylpolysiloxane XE 60 (25% cya-. noethyl, 75% methyl)
Good for aliphatic acid,carborane, acryl amide, organophosphorus pesticides, cocaine, phenols, aromatic amine. Temperature Limits: 0 to 250°C.
⑦ OV-1701 : 14% cyanopropylphenyl-86% methyl polysiloxane. alternative to HP-1701,RTX-1701,SPB-7,SPB-1701,007-1701,good for environmental, food and beverages, pesticides, herbicides. Temperature Limits: 0 to 300 °C.
Reference: Phenomenex, Agilent, Restek Cross Reference for Similar GC Columns.
Note: Next time, pls enclose more detail of GC assay parameters, being a good questioner will help us answer well, NOT just guess. Thanks for your question, Best.
How to improve this Deltamethrin GC chromatogram?
(YJ007, thanks for your question, highly appreciate your trust.)
Column: HP-5.
Oven Initial Temp: 150℃ hold 2min, then to 270℃@ 10℃/min for 23min.
Injection: Splitless @ 280, 1 ul.
Carrier gas: Helium @ 6ml/min.
ECD Temp: 300.
Make-Up gas: 30ml/min.
Deltamethrin Conc.: 0.1mg/L.
My answer to YJ007:
From this first chromatogram, the second peak in the 1st chromatogram should be the peak of Deltamethrin. However, that is NOT good one.
My first doubt was drawn by that flow rate (6ml/min): what is the diameter of that HP-5 column? Maybe 0.32 mm ( correct me if I am wrong) , so why use so high flow rate?
Any specific reason?
Tips: Increase the ECD Temp to 300 degree C, low down the initial temp to 110 degree C, adjust to flow rate according the diameter of the column. That maybe helps.
Here is my method used @ Feb-2005:
Column : HP-1, 5 m X 0.53 mm X 2.5 um
Inlet Temp: 280 ℃
ECD: 300 ℃
N2: purity NMT 99.99%, 13 ml/min
Then the retention time of Deltamethrin peak is 7.3 min.
Column: HP-5.
Oven Initial Temp: 150℃ hold 2min, then to 270℃@ 10℃/min for 23min.
Injection: Splitless @ 280, 1 ul.
Carrier gas: Helium @ 6ml/min.
ECD Temp: 300.
Make-Up gas: 30ml/min.
Deltamethrin Conc.: 0.1mg/L.
My answer to YJ007:
From this first chromatogram, the second peak in the 1st chromatogram should be the peak of Deltamethrin. However, that is NOT good one.
My first doubt was drawn by that flow rate (6ml/min): what is the diameter of that HP-5 column? Maybe 0.32 mm ( correct me if I am wrong) , so why use so high flow rate?
Any specific reason?
Tips: Increase the ECD Temp to 300 degree C, low down the initial temp to 110 degree C, adjust to flow rate according the diameter of the column. That maybe helps.
Here is my method used @ Feb-2005:
Column : HP-1, 5 m X 0.53 mm X 2.5 um
Inlet Temp: 280 ℃
ECD: 300 ℃
N2: purity NMT 99.99%, 13 ml/min
Then the retention time of Deltamethrin peak is 7.3 min.
Monday, July 21, 2008
Any Versatile Tool for Drug Residu Analysis?
Today, had a talked with Co-Worker "Is there any versatile Instrument for Residues Analysis ?"
As far as I know, the answer is NO for the residues analysis of following veterinary drug.
GC-MS: Terbutaline、Clenbuterol、Salbutamol、Ractopamine ( Clenbuterol , Clenbuterol, Spiropent; Planipart)、Cimaterol.
GC: Chloramphenicol,Chloromycetin and Polychlorinated biphenyl (PCBs)
LC-MS: Chlortetracycline,Aureomycin、Sulfonamides(SAs)、Ciprofloxacin、Enrofloxacin、clopidol、Tetracycline Tablets、Furazolidone、Olaquindox、Sudan yellow( oil orange), Stilbestrol.
One more phobia for the food:" why this chemicals matters? Just want to improve the meat products, NOT thinking about the side effect for our generations? Maybe is is time to re-evaluate this additives now, as least for me, thinking more greener if we could.
As far as I know, the answer is NO for the residues analysis of following veterinary drug.
GC-MS: Terbutaline、Clenbuterol、Salbutamol、Ractopamine ( Clenbuterol , Clenbuterol, Spiropent; Planipart)、Cimaterol.
GC: Chloramphenicol,Chloromycetin and Polychlorinated biphenyl (PCBs)
LC-MS: Chlortetracycline,Aureomycin、Sulfonamides(SAs)、Ciprofloxacin、Enrofloxacin、clopidol、Tetracycline Tablets、Furazolidone、Olaquindox、Sudan yellow( oil orange), Stilbestrol.
One more phobia for the food:" why this chemicals matters? Just want to improve the meat products, NOT thinking about the side effect for our generations? Maybe is is time to re-evaluate this additives now, as least for me, thinking more greener if we could.
Sunday, July 20, 2008
5973 Mass Spec failed to Pump (2)
Remove the side cover of vacuum room: pull out 70 C, hold firmly, lift it slowly.Remove the vacuum room and pump:Agilent always took it apart separately But for me, recommemed remove them together by un-screw the six screws
Remove the slot and vacuum line on the pump: anti-clickwise to take it downAsk Co-worker for help : hold the whole vacuum room first, to pull out the turbo pumpReplace the new end-cap, careful put them back. Enclose a back view of all the connections.
<See (1) at May-16-2008>
Subscribe to:
Posts (Atom)