Pesticides Analysis of 1,2-ethylene dibromide (EDB) and 1,2-dibromo-3-chloropropane (DMCP) by (Agilent 6890 Micro-ECDs) after microextraction with hexane in acc ordance with U.S. EPA method 504.
Relative standard deviation (% RSD) for the entire method was less than 7% over a concentration range greater than two orders of magnitude with method detection limits of 0.003 ug/L or lower. So Nice!
Assay Details:
Inj. Vol.: 2-mL splitless injection Inlet Split/splitless; 200 °C, pulsed splitless mode (20 psi for 1 min)
Carrier Helium: 6 psi (40 °C); 3.5 mL/min constant flow.
Column: (A) 30 m, 0.53-mm id, 0.8-mm film DB-608, an equivalent of HP-608. (B) 30 m, 0.53-mm id, 1.0-mm film RTX-1701, an equivalent of HP-PAS 1701.
Oven Temp: 40 °C (4 min); 10 °C/min to 240 °C;
Detector: 330 °C.
Makeup Gas: nitrogen, constant column and makeup flow (60 mL/min)