Problem: CHEMStation could not connect the MSD

Check the computer, power of MSD, could not fix the problem, even reinstall the ChemStation Software, the problem remains...
Oh, the problem maybe be lie in IO-auto config

Here are some steps how to fix that:
(1) go to program, then agilent IO libraries-------IO config

(2) Follow the instructions, highlight the ASRL1、USB0, then select "romove", Only leave the TCPIP0.

That is it......